MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,01,06,07 report with the FDA on 2014-04-08 for OXIMAX N-560 PULSE OXIMETER N560 manufactured by Mediana.
[17932203]
It was reported to covidien that part of the heart rate reading was lacking. No patient harm reported.
Patient Sequence No: 1, Text Type: D, B5
[18233685]
The circuit that displays the segments had a connection failure. The front board was replaced and the problem resolved. (b)(4). The serial number of the display pcb precedes a corrective and preventive action taken by the manufacturer.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 2936999-2014-00295 |
MDR Report Key | 3880249 |
Report Source | 00,01,06,07 |
Date Received | 2014-04-08 |
Date of Report | 2014-03-10 |
Date of Event | 2014-03-06 |
Date Mfgr Received | 2014-03-10 |
Device Manufacturer Date | 2013-03-01 |
Date Added to Maude | 2014-07-02 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | DENISE BRAXTON |
Manufacturer Street | 6135 GUNBARREL AVE. |
Manufacturer City | BOULDER CO 80301 |
Manufacturer Country | US |
Manufacturer Postal | 80301 |
Manufacturer Phone | 3038768909 |
Manufacturer G1 | MEDIANA |
Manufacturer Street | WONJU INDUSTRY PARK DONGWHA-RI, MUNMAK-EUP, WONJU-SI |
Manufacturer City | GANGWON-DO |
Manufacturer Country | KS |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | OXIMAX N-560 PULSE OXIMETER |
Generic Name | PULSE OXIMETER |
Product Code | DPZ |
Date Received | 2014-04-08 |
Returned To Mfg | 2014-03-18 |
Model Number | N-560 |
Catalog Number | N560 |
Device Availability | Y |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MEDIANA |
Manufacturer Address | WONJU INDUSTRY PARK DONGWHA-RI, MUNMAK-EUP GANGWON KS |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2014-04-08 |