ARCHITECT I2000SR ANALYZER 03M74-01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05 report with the FDA on 2014-06-26 for ARCHITECT I2000SR ANALYZER 03M74-01 manufactured by Abbott Manufacturing Inc.

Event Text Entries

[18934404] The customer stated that a (b)(6) result of (b)(6) was generated. The sample was repeated multiple times with results in the range of (b)(6). The patient had been run on another analyzer with a (b)(6) result. There was no report of impact to patient management.
Patient Sequence No: 1, Text Type: D, B5


[19134203] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[32678757] A false negative anti hbc result was generated for one patient on the architect i2000sr analyzer. Field service performed troubleshooting and determined the probable cause was the wash buffer dispense at the wash zones was inadequate. Service replaced the valve manifold kits to resolve the issue. No further discrepant results were noted. Customer complaint data was reviewed and no adverse trends were identified. The architect system operations manual was reviewed and was found to adequately address the issue. The investigation did not identify a malfunction / deficiency.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1628664-2014-00124
MDR Report Key3897760
Report Source01,05
Date Received2014-06-26
Date of Report2014-06-02
Date Mfgr Received2014-07-03
Device Manufacturer Date2003-02-20
Date Added to Maude2014-07-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone847937-512
Manufacturer G1ABBOTT MANUFACTURING INC
Manufacturer Street1921 HURD DRIVE
Manufacturer CityIRVING TX 75038
Manufacturer CountryUS
Manufacturer Postal Code75038
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARCHITECT I2000SR ANALYZER
Generic NameAUTOMATED IMMUNOASSAY ANALYZER
Product CodeMZP
Date Received2014-06-26
Catalog Number03M74-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MANUFACTURING INC
Manufacturer Address1921 HURD DRIVE IRVING TX 75038 US 75038


Patients

Patient NumberTreatmentOutcomeDate
10 2014-06-26

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.