DIMENSION VISTA 1500

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2014-07-03 for DIMENSION VISTA 1500 manufactured by Siemens Healthcare Diagnostics Inc.

Event Text Entries

[20657473] A discordant, falsely elevated chloride (cl) result was obtained on one patient sample on a dimension vista 1500 instrument. The discordant result was not reported to the physician(s). The sample was repeated twice on the same instrument, resulting lower. It is unknown if the corrected result was reported to the physician(s). There are no reports of patient intervention or adverse health consequences due to the discordant, falsely elevated cl result.
Patient Sequence No: 1, Text Type: D, B5


[20959407] A siemens customer service engineer (cse) was dispatched to the customer site. After evaluation of the instrument and instrument data, the cse cleaned the system, replaced the pump tubing, replaced and aligned the integrated multisensor technology probe and aliquot probe and aligned the imt rotary valve. The cause of the discordant, falsely elevated cl result is unknown. The instrument is performing according to specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1226181-2014-00346
MDR Report Key3913457
Report Source01,05,06
Date Received2014-07-03
Date of Report2014-06-11
Date of Event2014-05-27
Date Mfgr Received2014-06-11
Device Manufacturer Date2012-11-27
Date Added to Maude2014-07-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCASSANDRA KOCSIS
Manufacturer Street511 BENEDICT AVE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242687
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC
Manufacturer Street101 SILVERMINE ROAD
Manufacturer CityBROOKFIELD CT 06804
Manufacturer CountryUS
Manufacturer Postal Code06804
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDIMENSION VISTA 1500
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeCGZ
Date Received2014-07-03
Model NumberDIMENSION VISTA 1500
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC
Manufacturer Address101 SILVERMINE ROAD BROOKFIELD CT 06804 US 06804


Patients

Patient NumberTreatmentOutcomeDate
10 2014-07-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.