7500 FAST DX REAL-TIME PCR INSTRUMENT 4406985

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2014-06-11 for 7500 FAST DX REAL-TIME PCR INSTRUMENT 4406985 manufactured by Life Technologies Holdings Pte Ltd.

Event Text Entries

[4588691] During the operational qualification of the 7500 fast dx real-time pcr instrument (catalog no. 4406985, serial no. (b)(4)), the customer and field service engineer observed popping sounds and a burning smell. No pt involvement reported. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[12122389] Implant and explant dates: device is an instrument and is not implanted/explanted. The instrument was repaired by replacing the main controller board. After repair the unit was tested and passed performance specifications. Device intended use: the applied biosystems 7500 fast x real-time pcr instrument with the sds software version 1. 4 is a real-time nucleic acid amplification and detection system that measures nucleic acid signals from reverse transcribed rna and converts them to comparative quantitative readouts using fluorescent detection of dual-labeled hydrolysis probes. The 7500 fast dx real-time pcr instrument is to be used only by technologists trained in laboratory techniques, procedures, and on use of the analyzer.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3003673482-2014-00016
MDR Report Key3915637
Report Source07
Date Received2014-06-11
Date of Report2014-05-12
Date of Event2014-05-12
Date Mfgr Received2014-05-12
Device Manufacturer Date2006-08-01
Date Added to Maude2014-07-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKELLI TANZELLA
Manufacturer Street3175 NSTALEY RD
Manufacturer CityGRAND ISLAND NY 14072
Manufacturer CountryUS
Manufacturer Postal14072
Manufacturer Phone7167743122
Manufacturer G1LIFE TECHNOLOGIES HOLDINGS PTE LTD
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name7500 FAST DX REAL-TIME PCR INSTRUMENT
Generic NameINSTRUMENTATION FOR CLINICAL MULTIPL
Product CodeNSU
Date Received2014-06-11
Catalog Number4406985
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerLIFE TECHNOLOGIES HOLDINGS PTE LTD
Manufacturer AddressSINGAPORE SN


Patients

Patient NumberTreatmentOutcomeDate
10 2014-06-11