ZOOM AP LAMP ZM2500

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 99 report with the FDA on 2014-07-09 for ZOOM AP LAMP ZM2500 manufactured by Discus Dental, Llc.

Event Text Entries

[4584680] Hold -sdaon 05/20/2014, chris from the dental clinic called about a patient that had an ulcer on the gingiva of the bottom jaw from using the small grey color retractor. They stated that the patient had a hard time getting the small retractor in as it was still too big for this patient. They also stated that the patient is in pain and has been rinsing with warm water and salt. Patient also complained of sensitively, the dental office gave the patient a cream.
Patient Sequence No: 1, Text Type: D, B5


[11914827] The investigation included review of batch history record for zm2666 zoom ch kit lot # 13336048 (chairside gel 22-3764 lot # 13350030). The review of the batch history record did not uncover any adverse finding. In additional, the retain sample for sku # 22-3764 lot #13350030 was tested and it was found to be within specifications. No other quality issue was revealed during the review of said records. The device history record for the zoom ap lamp, zm2500 serial # 02376 is not available for review since the lamp was manufactured in 2005 and the record retention for the dhr is five years. Discus dental will continue to monitor similar issues. See scanned pages.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1000582314-2014-00011
MDR Report Key3927368
Report Source99
Date Received2014-07-09
Date of Report2014-06-12
Date Mfgr Received2014-05-20
Device Manufacturer Date2005-05-01
Date Added to Maude2014-07-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCARMELITA TORRES
Manufacturer Street1700 A SOUTH BAKER AVE.
Manufacturer CityONTARIO CA 91761
Manufacturer CountryUS
Manufacturer Postal91761
Manufacturer Phone9095703512
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameZOOM AP LAMP
Generic NameWHITENING LAMP
Product CodeEEG
Date Received2014-07-09
Model NumberZM2500
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerDISCUS DENTAL, LLC
Manufacturer AddressONTARIO CA 91761 US 91761


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2014-07-09