MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2014-07-11 for VITROS CHEMISTRY PRODUCTS CRBM SLIDES 8892382 manufactured by Ortho-clinical Diagnostics.
[19719575]
The customer obtained non-reproducible, higher than expected quality control results on two vitros microslide assays tested on a vitros 5600 integrated system: vitros crbm: control l1: 8. 99 ug/ml vs expected 4. 5 ug/ml. Vitros phyt: control l1: 14. 29 ug/ml vs expected 7. 6 ug/ml. Biased results of the magnitude and direction observed may lead to inappropriate physician action if patient samples were affected. No patient samples were known to have been affected or reported from the laboratory. There was no allegation of patient harm as a result of this event. This report is number two of two mdr? S for this event. Two 3500a forms are being submitted for this event as two devices were involved. This report corresponds to ortho clinical diagnostics inc. , complaint number (b)(4).
Patient Sequence No: 1, Text Type: D, B5
[19858278]
The investigation determined that the non-reproducible, higher than expected quality control results were obtained on vitros crbm and phyt assays on a vitros 5600 integrated system. There was no evidence to suggest a malfunction of either vitros microslide reagents or vitros instrument involved. In addition, pre-analytical sample mix-up is a possible contributing factor, although this could not be confirmed. The assignable cause of this event is unknown.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1319809-2014-00029 |
MDR Report Key | 3929831 |
Report Source | 05 |
Date Received | 2014-07-11 |
Date of Report | 2014-07-11 |
Date of Event | 2014-05-21 |
Date Mfgr Received | 2014-06-12 |
Device Manufacturer Date | 2014-02-18 |
Date Added to Maude | 2014-07-24 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 0 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | MR. JAMES A STEVENS |
Manufacturer Street | 100 INDIGO CREEK DRIVE |
Manufacturer City | ROCHESTER NY 14626 |
Manufacturer Country | US |
Manufacturer Postal | 14626 |
Manufacturer Phone | 5854533000 |
Manufacturer G1 | ORTHO-CLINICAL DIAGNOSTICS |
Manufacturer Street | 513 TECHNOLOGY BLVD. |
Manufacturer City | ROCHESTER NY 14652 |
Manufacturer Country | US |
Manufacturer Postal Code | 14652 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | VITROS CHEMISTRY PRODUCTS CRBM SLIDES |
Generic Name | IN-VITRO DIAGNOSTIC |
Product Code | KLT |
Date Received | 2014-07-11 |
Catalog Number | 8892382 |
Lot Number | 3921-0091-1485 |
Device Expiration Date | 2014-11-01 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ORTHO-CLINICAL DIAGNOSTICS |
Manufacturer Address | 100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US 14626 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2014-07-11 |