VITROS CHEMISTRY PRODUCTS CRBM SLIDES 8892382

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2014-07-11 for VITROS CHEMISTRY PRODUCTS CRBM SLIDES 8892382 manufactured by Ortho-clinical Diagnostics.

Event Text Entries

[19719575] The customer obtained non-reproducible, higher than expected quality control results on two vitros microslide assays tested on a vitros 5600 integrated system: vitros crbm: control l1: 8. 99 ug/ml vs expected 4. 5 ug/ml. Vitros phyt: control l1: 14. 29 ug/ml vs expected 7. 6 ug/ml. Biased results of the magnitude and direction observed may lead to inappropriate physician action if patient samples were affected. No patient samples were known to have been affected or reported from the laboratory. There was no allegation of patient harm as a result of this event. This report is number two of two mdr? S for this event. Two 3500a forms are being submitted for this event as two devices were involved. This report corresponds to ortho clinical diagnostics inc. , complaint number (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[19858278] The investigation determined that the non-reproducible, higher than expected quality control results were obtained on vitros crbm and phyt assays on a vitros 5600 integrated system. There was no evidence to suggest a malfunction of either vitros microslide reagents or vitros instrument involved. In addition, pre-analytical sample mix-up is a possible contributing factor, although this could not be confirmed. The assignable cause of this event is unknown.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1319809-2014-00029
MDR Report Key3929831
Report Source05
Date Received2014-07-11
Date of Report2014-07-11
Date of Event2014-05-21
Date Mfgr Received2014-06-12
Device Manufacturer Date2014-02-18
Date Added to Maude2014-07-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. JAMES A STEVENS
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal14626
Manufacturer Phone5854533000
Manufacturer G1ORTHO-CLINICAL DIAGNOSTICS
Manufacturer Street513 TECHNOLOGY BLVD.
Manufacturer CityROCHESTER NY 14652
Manufacturer CountryUS
Manufacturer Postal Code14652
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVITROS CHEMISTRY PRODUCTS CRBM SLIDES
Generic NameIN-VITRO DIAGNOSTIC
Product CodeKLT
Date Received2014-07-11
Catalog Number8892382
Lot Number3921-0091-1485
Device Expiration Date2014-11-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US 14626


Patients

Patient NumberTreatmentOutcomeDate
10 2014-07-11

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.