MICROFRANCE? INSTRUMENT MCO13C

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,07 report with the FDA on 2014-07-14 for MICROFRANCE? INSTRUMENT MCO13C manufactured by Xomed Microfrance Mfg.

Event Text Entries

[18039872] The device was returned with a request for repair. Analysis found that the mobile jaw of the instrument is missing. There was no reported patient impact.
Patient Sequence No: 1, Text Type: D, B5


[18130973] (b)(4): product evaluation: analysis found that the mobile jaw of the instrument is missing. The fragment was not returned by the client. Considering the size of the fragment (1cm) and the type of surgery (ear microsurgery), the fragment is probably not in patient contact. The fixed jaw is badly damaged. The pin between the body and the mobile jaw is broken and its fragment is still on the forceps. The most probable cause is an excessive effort during the use or the reprocessing of the instrument. Results: fracture problem. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9680837-2014-00056
MDR Report Key3932078
Report Source01,07
Date Received2014-07-14
Date of Report2014-06-19
Date of Event2014-06-12
Date Mfgr Received2014-06-19
Device Manufacturer Date2014-02-01
Date Added to Maude2014-07-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMICHELLE ALFORD
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal32216
Manufacturer Phone9043328197
Manufacturer G1MEDTRONIC XOMED, INC.
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal Code32216
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMICROFRANCE? INSTRUMENT
Generic NameFORCEPS, ENT
Product CodeKAE
Date Received2014-07-14
Returned To Mfg2014-06-19
Model NumberMCO13C
Catalog NumberMCO13C
Lot Number140203
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerXOMED MICROFRANCE MFG
Manufacturer AddressSAINT-AUBIN-LE-MONIAL BOURBON-L'ARCHAMBAULT 3160 FR 3160


Patients

Patient NumberTreatmentOutcomeDate
10 2014-07-14