MICROFRANCE? INSTRUMENT MCO205A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06,07 report with the FDA on 2014-07-17 for MICROFRANCE? INSTRUMENT MCO205A manufactured by Xomed Microfrance Mfg.

Event Text Entries

[4572079] The device came in for service with the comment? Already broken.? The product analysis indicated that the tip of the hook was broken and the missing fragment was not returned.
Patient Sequence No: 1, Text Type: D, B5


[12159538] Concomitant hook mco205b 0. 3mm 90deg anterior, 510k: exempt, lot# 140203, mfr date feb. 2014 (b)(4). The product analysis determined that for both the anterior and posterior hooks (hook mco205a 0. 3mm 90deg posterior and hook mco205b 0. 3mm 90deg anterior),? The tip of the hook is broken. The missing fragment was not returned. Considering its size (approximatively 0. 1mm), the patient risk is very low. Moreover, it appears that the event happened during the reprocessing of the instrument. The distal part of the instrument is bent. No manufacturing or material defect was found. The most probable cause of this breakage is an excessive effort during the use or the reprocessing of the instrument.? (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9680837-2014-00058
MDR Report Key3941282
Report Source01,05,06,07
Date Received2014-07-17
Date of Report2014-06-24
Date of Event2014-06-19
Date Mfgr Received2014-06-24
Device Manufacturer Date2014-02-01
Date Added to Maude2014-07-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactAMY CORRALES
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal32216
Manufacturer Phone9043328138
Manufacturer G1MEDTRONIC XOMED, INC.
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal Code32216
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMICROFRANCE? INSTRUMENT
Generic NameHOOK, MICROSURGICAL EAR
Product CodeJYL
Date Received2014-07-17
Returned To Mfg2014-06-24
Model NumberMCO205A
Catalog NumberMCO205A
Lot Number140203
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerXOMED MICROFRANCE MFG
Manufacturer AddressSAINT-AUBIN-LE-MONIAL BOURBON-L'ARCHAMBAULT 3160 FR 3160


Patients

Patient NumberTreatmentOutcomeDate
10 2014-07-17