MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2014-07-18 for STERNUM BLADE 0277087100 manufactured by Stryker Instruments-kalamazoo.
[5015444]
It was reported that during a procedure the saw blade broke during the cut. When the blade break was noticed an x-ray was taken and the other part of the blade was found in the patient. It was further reported that this resulted in a surgical delay of 15-20 minutes. The procedure was completed successfully.
Patient Sequence No: 1, Text Type: D, B5
[12557176]
The device was not returned for evaluation. The quality investigation is complete. Device not returned to the manufacturer.
Patient Sequence No: 1, Text Type: N, H10
[21488964]
It was reported that during a procedure the saw blade broke during the cut. When the blade break was noticed, an x-ray was taken and the other part of the blade was found in the patient. It was further reported that this resulted in a surgical delay of 15-20 minutes. The procedure was completed successfully.
Patient Sequence No: 1, Text Type: D, B5
[21737204]
Device return has been requested. A follow up report will be filed once the quality investigation is complete. Device not yet returned to manufacturer.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 0001811755-2014-02573 |
MDR Report Key | 3943781 |
Report Source | 06 |
Date Received | 2014-07-18 |
Date of Report | 2014-06-24 |
Date of Event | 2014-06-24 |
Date Mfgr Received | 2014-11-11 |
Date Added to Maude | 2014-07-18 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | MR. DERVILLIA MURPHY |
Manufacturer Street | INSTRUMENTS DIVISION CARRIGTWOHILL BUS. & TECH PARK |
Manufacturer City | CARRIGTWOHILL NA |
Manufacturer Postal | NA |
Manufacturer Phone | 214532900 |
Manufacturer G1 | STRYKER INSTRUMENTS-IRELAND |
Manufacturer Street | INSTRUMENTS DIVISION CARRIGTWOHILL BUS. & TECH PARK |
Manufacturer City | CARRIGTWOHILL NA |
Manufacturer Postal Code | NA |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | STERNUM BLADE |
Generic Name | BLADE, SAW, SURGICAL, CARDIOVASCULAR |
Product Code | DWH |
Date Received | 2014-07-18 |
Catalog Number | 0277087100 |
Lot Number | UNKNOWN |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | STRYKER INSTRUMENTS-KALAMAZOO |
Manufacturer Address | 4100 EAST MILHAM AVENUE KALAMAZOO MI 49001 US 49001 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2014-07-18 |