REOCOR D 365529 SEE MODEL NO.

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,07 report with the FDA on 2014-07-21 for REOCOR D 365529 SEE MODEL NO. manufactured by Biotronik Se & Co. Kg.

Event Text Entries

[4645983] Ous mdr - it was reported that the device is not capturing properly. No details available at the moment. No adverse patient side effects were reported.
Patient Sequence No: 1, Text Type: D, B5


[17813015] Upon receipt, the external pacemaker was inspected. Despite the damaged case, which was repaired, the device did not show any deviation during analysis. No indication which might have contributed to the clinical observation was revealed. The visual, mechanical and functional analysis revealed no anomalies and no material or manufacturing deviation was found.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1028232-2014-002440
MDR Report Key3947446
Report Source01,07
Date Received2014-07-21
Date of Report2014-07-08
Date Mfgr Received2014-08-04
Device Manufacturer Date2010-10-05
Date Added to Maude2014-07-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street6024 JEAN ROAD
Manufacturer CityLAKE OSWEGO OR 97035
Manufacturer CountryUS
Manufacturer Postal97035
Manufacturer Phone8772459800
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameREOCOR D
Generic NameEXTERNAL PACEMAKER
Product CodeOVJ
Date Received2014-07-21
Model Number365529
Catalog NumberSEE MODEL NO.
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBIOTRONIK SE & CO. KG
Manufacturer AddressWOERMANNKEHRE 1 BERLIN D-12359 GM D-12359


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2014-07-21

© 2026 FDA.report