[4642919] A customer reported a software bug which resulted in incorrect sample identification. When one sample on a plate was assigned a new identifier, the rest of the samples on the plate were automatically renamed. Additionally, a pop-up error message appeared on the screen. The medical device involved was quantstudio dx real-time pcr instrument (catalog no. 4484383, serial no. (b)(4)). No patient involvement reported. The company event tracking no. Is (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[11982134] Implant and explant dates: device is an instrument and is not implanted/explanted. The instrument was investigated on site. The root cause was determined to be a software design issue which prevented the customer from obtaining results after a run. To by-pass the software issue, the field application specialist instructed the customer on how to use the "import from file" function in the software. This allowed the customer to obtain results after the run. The software issue has been addressed and a new version of the software is scheduled for release.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3003673482-2014-00009
MDR Report Key3949740
Report Source04,07
Date Received2014-05-02
Date of Report2013-10-24
Date of Event2013-10-24
Date Mfgr Received2013-10-24
Device Manufacturer Date2013-03-01
Date Added to Maude2014-07-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKELLI TANZELLA
Manufacturer Street3175 STALEY ROAD
Manufacturer CityGRAND ISLAND NY 14072
Manufacturer CountryUS
Manufacturer Postal14072
Manufacturer Phone7167743122
Manufacturer G1LIFE TECHNOLOGIES HOLDINGS PTE. LTD.
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameQUANTSTUDIO DX REAL-TIME PCR INSTRUMENT
Generic NameREAL TIME NUCLEIC ACID AMPLIFICATION
Product CodeOOI
Date Received2014-05-02
Catalog Number4484383
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerLIFE TECHNOLOGIES HOLDINGS PTE. LTD.
Manufacturer AddressSN


Patients

Patient NumberTreatmentOutcomeDate
10 2014-05-02