HEARTMATE RECHARGEABLE BATTERY 102515

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2014-06-13 for HEARTMATE RECHARGEABLE BATTERY 102515 manufactured by Thoratec Corp..

Event Text Entries

[4644605] The patient was implanted with a left ventricular assist device (lvad). It was reported by the physician that the patient went to the emergency room due to a message "clock not set" displayed on his system controller. The patient stated that he fell asleep the night before a approximately 11:00pm and did not wake up until 1:00pm the next day. When the patient woke up, he stated that the batteries were dead and the pump was "off". He exchanged the batteries and the pump started. When interrogating the history, the physician saw "low voltage" advisories starting at 4:02am. The patient was asymptomatic. The physician was instructed on resyncing the date/time to the system controller and the alarm cleared. It was also suggested by the clinical consultant that a heparin gtts be started due to the undetermined amount of time the pump may have been off.
Patient Sequence No: 1, Text Type: D, B5


[11975975] The manufacturer is attempting to acquire the device for further evaluation. No further information is available at this time. A supplemental report will be submitted when the manufacturer's investigation is completed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2916596-2014-00993
MDR Report Key3950795
Report Source05,06
Date Received2014-06-13
Date of Report2014-05-17
Date of Event2014-05-17
Date Facility Aware2014-05-17
Report Date2014-05-17
Date Reported to Mfgr2014-05-17
Date Mfgr Received2014-05-17
Date Added to Maude2014-07-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. ROBERT FRYC
Manufacturer Street23 FOURTH AVENUE
Manufacturer CityBURLINGTON MA 01803
Manufacturer CountryUS
Manufacturer Postal01803
Manufacturer Phone7812720139
Manufacturer G1THORATEC CORP.
Manufacturer Street6035 STONERIDGE DR.
Manufacturer CityPLEASANTON CA 94588
Manufacturer CountryUS
Manufacturer Postal Code94588
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHEARTMATE RECHARGEABLE BATTERY
Generic NameRECHARGEABLE BATTERY
Product CodeMOY
Date Received2014-06-13
Model Number102515
Catalog Number102515
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Age3 MO
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerTHORATEC CORP.
Manufacturer AddressPLEASANTON CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2014-06-13