SCALE-TRONIX 4002 INFANT SCALE 42110

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2014-07-18 for SCALE-TRONIX 4002 INFANT SCALE 42110 manufactured by Scale-tronix, Inc..

Event Text Entries

[4623855] While doing preventive maintenance(pm) on the ohmeda 4400 open infant warmer and checking the in-bed scale-tronix 4002 scale, the biomed tech noticed a potential safety hazard. The scale underneath each mattress has two pieces, the top piece and the actual scale. They clip together with two silver spring-loaded clips/pins. During the pm it was found that the two pieces of the scale would not completely clip together and as a result the top piece of the scale was slipping off the scale. The hole for the spring clip was worn and an edge developed so the clip was not engaging properly. The mattress sits on the upper piece of the scale. Under ideal circumstances if the bed "wing" was left down and if these two scale pieces are not clipped together properly the baby, mattress, and top section of scale have the potential to slide out and fall. The problem seems to come from normal wear and tear on these scales. We are reporting this as something to watch for during maintenance. We have advised our nursing staff of the potential risk with this issue. In the meantime we are inspecting all of our beds for this problem.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3956937
MDR Report Key3956937
Date Received2014-07-18
Date of Report2014-07-18
Date of Event2014-07-11
Report Date2014-07-18
Date Reported to FDA2014-07-18
Date Reported to Mfgr2014-07-25
Date Added to Maude2014-07-25
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameSCALE-TRONIX 4002 INFANT SCALE
Generic NameSCALE, PATIENT
Product CodeFRW
Date Received2014-07-18
Model Number4002
Catalog Number42110
Lot Number*
ID Number*
Device AvailabilityY
Device Age*
Device Sequence No1
Device Event Key0
ManufacturerSCALE-TRONIX, INC.
Manufacturer Address288 CARLTON DR CAROL STREAM IL 60188 US 60188


Patients

Patient NumberTreatmentOutcomeDate
10 2014-07-18

© 2024 FDA.report