A CORD MH-969 NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2014-06-13 for A CORD MH-969 NA manufactured by Olympus Medical Systems Corporation.

Event Text Entries

[4556638] Olympus medical systems corp (omsc) was informed that during polypectomy of colon the doctor got an error trying to activate output while he snared the target polyp and output could not be activated. Then he noticed that the subject product broke. He exchanged the subject product and the electrosurgical unit for other devices and completed the procedure. There was no repair of patient injury regarding this report.
Patient Sequence No: 1, Text Type: D, B5


[12099156] The subject device was returned to omsc for investigation. The investigation confirmed that the cable broke near a code plug. The cable core was burnt and the broken out section was discolored. Omsc was informed that the user facility had used the subject device for nearly 10 years. Omsc assumes that bending load was put on the cable repeatedly over long periods and it caused the breakage of the cable core. In addition, the user facility continued to use the subject device with the cable core broken, it caused this event. This report is being submitted as a medical devic report in an abundance of caution.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8010047-2014-00310
MDR Report Key3958438
Report Source05,06
Date Received2014-06-13
Date of Report2014-05-19
Date of Event2014-05-16
Date Mfgr Received2014-05-19
Date Added to Maude2014-07-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactHIROKI MORIYAMA
Manufacturer Street2951 ISHIKAWA-CHO
Manufacturer CityHACHIOJI-SHI, TOKYO 192-8507
Manufacturer CountryJA
Manufacturer Postal192-8507
Manufacturer Phone26425177
Single Use0
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameA CORD
Generic NameA CORD
Product CodeFFZ
Date Received2014-06-13
Returned To Mfg2014-05-23
Model NumberMH-969
Catalog NumberNA
Lot NumberUNK
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS MEDICAL SYSTEMS CORPORATION
Manufacturer Address2951 ISHIKAWA-CHO HACHIOJI-SHI, TOKYO 12-8507 JA 12-8507