FOLEY CATHETER, 16FR SILICONE DYND160816

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2014-05-23 for FOLEY CATHETER, 16FR SILICONE DYND160816 manufactured by Medline Industries, Inc..

Event Text Entries

[4625530] The foley catheter fell out and the balloon was found to be ruptured.
Patient Sequence No: 1, Text Type: D, B5


[11939646] The catheter fell out and the balloon was found to be ruptured. There was no pt injury and a new catheter was not inserted. No medical intervention was indicated. The sample was returned and evaluated. The balloon had a clean tear. It is not known how much fluid was used to inflate the balloon. We cannot rule out over-inflation of the catheter balloon as a potential root cause for the rupture.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1417592-2014-00054
MDR Report Key3961011
Report Source06
Date Received2014-05-23
Date of Report2014-05-22
Date of Event2014-04-22
Date Mfgr Received2014-04-23
Device Manufacturer Date2013-11-01
Date Added to Maude2014-08-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactJULIE FINLEY
Manufacturer StreetONE MEDLINE PLACE
Manufacturer CityMUNDELEIN IL 60060
Manufacturer CountryUS
Manufacturer Postal60060
Manufacturer Phone8476434709
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameFOLEY CATHETER, 16FR SILICONE
Product CodeNWR
Date Received2014-05-23
Catalog NumberDYND160816
Lot Number13WB3883
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDLINE INDUSTRIES, INC.
Manufacturer AddressMUNDELEIN IL 60060 US 60060


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2014-05-23

© 2026 FDA.report