OXIMAX N-560 PULSE OXIMETER N560

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,01,06 report with the FDA on 2014-05-30 for OXIMAX N-560 PULSE OXIMETER N560 manufactured by Mediana.

Event Text Entries

[4556417] Covidien received a report of a n560 for a missing segment of the display. The failure happened when the n560 was not being used on a patient, the technician was checking the device.
Patient Sequence No: 1, Text Type: D, B5


[12060269] (b)(4). Covidien service center verified and isolated the failure to the front pcb board. Manufacturing date of this unit precedes a corrective and preventive action taken by the manufacturer for this failure mode.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2936999-2014-00432
MDR Report Key3964596
Report Source00,01,06
Date Received2014-05-30
Date of Report2014-05-08
Date of Event2014-05-01
Date Mfgr Received2014-05-08
Device Manufacturer Date2013-02-01
Date Added to Maude2014-08-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactDENISE BRAXTON
Manufacturer Street6135 GUNBARREL AVE.
Manufacturer CityBOULDER CO 80301
Manufacturer CountryUS
Manufacturer Postal80301
Manufacturer Phone3038768909
Manufacturer G1MEDIANA
Manufacturer StreetWONJU MEDICAL INDUSTRY PARK DONGHWA-RI, MUNMAK-EUP, WONJU-SI
Manufacturer CityGANGWON-DO
Manufacturer CountryKS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOXIMAX N-560 PULSE OXIMETER
Generic NamePULSE OXIMETER
Product CodeDPZ
Date Received2014-05-30
Returned To Mfg2014-05-08
Model NumberN-560
Catalog NumberN560
OperatorOTHER
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDIANA
Manufacturer AddressWONJU INDUSTRY DONGHWA-RI, MUNMAK-EUP GANGWON KS


Patients

Patient NumberTreatmentOutcomeDate
10 2014-05-30

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.