DADE ACTIN FSL ACTIVATED PTT REAGENT B4219-2

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2014-07-30 for DADE ACTIN FSL ACTIVATED PTT REAGENT B4219-2 manufactured by Siemens Healthcare Diagnostics Products Gmbh.

Event Text Entries

[4609104] Depressed activated partial thromboplastin time (aptt) results well below laboratory reference ranges were obtained on patient samples. Patient results were reported to the physician who questioned the results as discordant low. The account notes that there was a delay in patient surgery due to the repeat testing of the depressed aptt results for one patient. It is unknown if patient treatment was otherwise altered or prescribed due to questioned, depressed aptt results. There is no report of adverse outcome to the patients as a result of the questioned, depressed aptt results.
Patient Sequence No: 1, Text Type: D, B5


[12150518] The cause of the discrepant depressed activated partial thromboplastin time results is unknown. Siemens healthcare diagnostics has neither an indication for a malfunction of the analyzer nor for a performance issue of the reagent being used. The instrument is performing within specifications.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610806-2014-00021
MDR Report Key3965825
Report Source05,06
Date Received2014-07-30
Date of Report2014-06-30
Date of Event2014-06-29
Date Mfgr Received2014-06-30
Device Manufacturer Date2013-05-08
Date Added to Maude2014-08-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactAYDEE CRAWFORD
Manufacturer StreetGLASGOW BUSINESS COMMUNITY PO BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal197146101
Manufacturer Phone3026316312
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer Street76 EMIL VON BEHRING STRASSE
Manufacturer CityMARBURG, 35041
Manufacturer CountryGM
Manufacturer Postal Code35041
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDADE ACTIN FSL ACTIVATED PTT REAGENT
Generic NameDADE ACTIN FSL ACTIVATED PTT REAGENT
Product CodeGGW
Date Received2014-07-30
Catalog NumberB4219-2
Lot Number547389
Device Expiration Date2015-10-09
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer Address76 EMIL VON BEHRING STRASSE MARBURG, 35041 GM 35041


Patients

Patient NumberTreatmentOutcomeDate
10 2014-07-30