[275527] Customer reported a washer module lockup with ortho summit processor and that the last 3 rows of an ortho hcv 3. 0 plate were not washed. The instrument did not invalidate the wells. Test results for these wells would have been reactive, since the conjugate was not removed from the wells. An fse was dispatched and the washer manifold and boards were replaced. Diagnostic checks were performed to verify that the instrument was operating properly after repair. No death or serious injury was associated with this incident.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2250051-2002-00569
MDR Report Key397302
Date Received2002-05-16
Date of Report2002-05-14
Date of Event2002-03-28
Date Facility Aware2002-04-01
Report Date2002-04-06
Date Reported to Mfgr2002-05-14
Date Added to Maude2002-06-04
Event Key0
Report Source CodeDistributor report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameORTHO SUMMIT PROCESSOR
Generic NameSAMPLE PROCESSOR
Product CodeJTL
Date Received2002-05-16
Model Number7003015
Catalog Number936480
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age4 YR
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key386355
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address1001 US HWY 2002 RARITAN NJ 088690606 US
Baseline Brand NameORTHO SUMMIT PROCESSOR
Baseline Generic NameSAMPLE PROCESSOR
Baseline Model No7003015
Baseline Catalog No936480
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2002-05-16