WRO 300H

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2014-08-01 for WRO 300H manufactured by Mar Cor Purification.

Event Text Entries

[4608700] Customer contacted mar cor (mcp) to report their portable reverse osmosis machines were leaking with water going into adjacent rooms. Patient treatment delayed.
Patient Sequence No: 1, Text Type: D, B5


[12072443] The customer stated that some treatments were delayed due to the leaking. Replacement valves were shipped for all 11 machines because customer could not identify which machines were leaking. The leaking part(s) was a backflow preventer on the incoming supply water to the portable ro machines. This backflow preventer is not part of the device but was installed on their incoming water lines at the customer's request to meet local plumbing code requirements. Measurement of the incoming water supply showed wide pressure fluctuations from 20psi to 60 psi on a routine basis. A backflow preventer specialist is evaluating to identify a different valve to utilize. This will be monitored within marcor purification complaint system. If additional information is received, this complaint will be evaluated.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3019131-2014-00024
MDR Report Key3973173
Report Source05,06
Date Received2014-08-01
Date of Report2014-08-01
Date of Event2014-07-03
Date Mfgr Received2014-07-03
Date Added to Maude2014-08-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMEGAN DICKEY
Manufacturer Street14550 28TH AVE N
Manufacturer CityMINNEAPOLIS MN 55447
Manufacturer CountryUS
Manufacturer Postal55447
Manufacturer Phone7635533300
Manufacturer G1MAR COR PURIFICATION
Manufacturer Street14550 28TH AVE N
Manufacturer CityMINNEAPOLIS MN 55447
Manufacturer CountryUS
Manufacturer Postal Code55447
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWRO 300H
Generic NameWATER PURIFICATION SYSTEM FOR HEMODIALYSIS
Product CodeFIP
Date Received2014-08-01
OperatorOTHER
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMAR COR PURIFICATION
Manufacturer Address14550 28TH AVE N MINNEAPOLIS MN 55447 US 55447


Patients

Patient NumberTreatmentOutcomeDate
10 2014-08-01

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.