APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR 4407205

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04,07 report with the FDA on 2014-06-27 for APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR 4407205 manufactured by Life Technologies Holdings Pte Ltd.

Event Text Entries

[4822769] A (b)(6) field svc engineer reported that contamination from the heated cover of applied biosystems 7500 fast dx real-time pcr instrument (catalog no. 4407205 serial no. (b)(4)) caused incorrect results to be reported from a run. We were not made aware of any impact to pt management. The company event tracking no: (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[11943802] Device is an instrument and is not implanted/explanted. The field svc engineer attributed the contamination issue to wear and tear of the heated cover. The heated cover was replaced and the contamination issue was resolved. Device intended use: the applied biosystems 7500 fast dx real-time pcr instrument with the sds software version 1. 4 is a real-time nucleic acid amplification and detection system that measures nucleic acid signals from reverse transcribed rna and converts them to comparative quantitative readouts using fluorescent detection of dual-labeled hydrolysis probes. The 7500 fast dx real-time pcr instrument is to be used only by technologists trained in laboratory techniques, procedures and on use of the analyzer.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3003673482-2014-00018
MDR Report Key3978888
Report Source04,07
Date Received2014-06-27
Date of Report2013-03-07
Date of Event2013-03-07
Date Mfgr Received2013-03-07
Device Manufacturer Date2009-09-01
Date Added to Maude2014-08-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKELLI TANZELLA
Manufacturer Street3175 STALEY RD
Manufacturer CityGRAND ISLAND NY 14072
Manufacturer CountryUS
Manufacturer Postal14072
Manufacturer Phone7167743122
Manufacturer G1LIFE TECHNOLOGIES HOLDINGS PTE LTD
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAPPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR
Generic NameINSTRUMENTATION FOR CLINICAL MULTIP
Product CodeNSU
Date Received2014-06-27
Catalog Number4407205
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerLIFE TECHNOLOGIES HOLDINGS PTE LTD
Manufacturer AddressSINGAPORE SN


Patients

Patient NumberTreatmentOutcomeDate
10 2014-06-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.