MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2014-08-05 for X7000 XENON LIGHTSOURCE 0220190000 manufactured by Stryker Endoscopy-san Jose.
[17616861]
It was reported that the lightsource automatically switched to standby intermittently. The procedure was later switched to an open surgery.
Patient Sequence No: 1, Text Type: D, B5
[17811748]
Additional information will be provided once the investigation has been completed.
Patient Sequence No: 1, Text Type: N, H10
[30273333]
X7000 s/n (b)(4) was returned to stryker for analysis, but the failure mode was not confirmed. The customer complaint of? Switched to standby? Could not be duplicated. Visual inspection: there were no outward signs of cosmetic or physical damage to the product. Functional inspection: all measurements and functionality testing met specification. Probable root cause: could not duplicate the complaint in our test environment. In sum, the unit was returned, but the failure mode was not confirmed.
Patient Sequence No: 1, Text Type: N, H10
[30273334]
It was reported that the lightsource automatically switched to standby intermittently. The procedure was later switched to an open surgery.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 0002936485-2014-00576 |
MDR Report Key | 3982333 |
Report Source | 07 |
Date Received | 2014-08-05 |
Date of Report | 2014-07-09 |
Date of Event | 2014-07-09 |
Date Mfgr Received | 2014-07-09 |
Date Added to Maude | 2014-08-14 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | MR. THOMAS SHAFER |
Manufacturer Street | 5900 OPTICAL COURT |
Manufacturer City | SAN JOSE CA 95138 |
Manufacturer Country | US |
Manufacturer Postal | 95138 |
Manufacturer Phone | 4087542000 |
Manufacturer G1 | STRYKER ENDOSCOPY-SAN JOSE |
Manufacturer Street | 5900 OPTICAL COURT |
Manufacturer City | SAN JOSE CA 95138 |
Manufacturer Country | US |
Manufacturer Postal Code | 95138 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | X7000 XENON LIGHTSOURCE |
Generic Name | LIGHT, SURGICAL, FLOOR STANDING |
Product Code | FSS |
Date Received | 2014-08-05 |
Returned To Mfg | 2014-07-28 |
Catalog Number | 0220190000 |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | STRYKER ENDOSCOPY-SAN JOSE |
Manufacturer Address | 5900 OPTICAL COURT SAN JOSE CA 95138 US 95138 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2014-08-05 |