ORCHIDE OHD SET 567702910C

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06,07 report with the FDA on 2014-06-11 for ORCHIDE OHD SET 567702910C manufactured by Maquet S.a.s..

Event Text Entries

[4885979] The hosp reported that a camera broke and detached from the light head while being manipulated by a doctor. The camera remained attached to the light head, suspended by its cables. No injuries nor adverse effects were reported. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[12323086] A maquet field service tech (fst) evaluated the device. He found evidence this device collided with other objects in the operating room. Repeated impacts of the device are capable of dislodging the camera. The fst replaced the camera with a new one. The orchide camera yearly maintenance program includes a verification of the condition of the plastic covers and absence of scratches. Maquet med systems usa submits this report on behalf of the device mfg facility. Maquet (b)(4) provides product failure investigation, analysis and resolution for the device described in this report.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9710055-2014-00043
MDR Report Key3987468
Report Source01,05,06,07
Date Received2014-06-11
Date of Report2014-05-16
Date Facility Aware2014-06-11
Report Date2014-06-11
Date Mfgr Received2014-05-16
Device Manufacturer Date2009-11-01
Date Added to Maude2014-08-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJANICE PEVIDE
Manufacturer Street45 BARBOUR POND DRIVE
Manufacturer CityWAYNE NJ 07470
Manufacturer CountryUS
Manufacturer Postal07470
Manufacturer Phone9737097753
Manufacturer G1MAQUET S.A.S
Manufacturer StreetPARC DE LIMERE AVENUE DE LA POMME DE PIN
Manufacturer CityARDON, ORLEANS CEDEX 2 45074
Manufacturer CountryFR
Manufacturer Postal Code45074
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameORCHIDE
Product CodeKQM
Date Received2014-06-11
Model NumberOHD SET
Catalog Number567702910C
Lot NumberNA
ID Number567702904
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age4 YR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMAQUET S.A.S.
Manufacturer AddressORLEANS FR


Patients

Patient NumberTreatmentOutcomeDate
10 2014-06-11

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.