STRYKERVISION FLAT-PANEL MOUNTING ARM 0240095200

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2014-08-11 for STRYKERVISION FLAT-PANEL MOUNTING ARM 0240095200 manufactured by Stryker Endoscopy-san Jose.

Event Text Entries

[4957554] It was reported that surgeon was hit on the shoulder and the nurse was hit in the face while trying to adjust the angle of the monitor.
Patient Sequence No: 1, Text Type: D, B5


[11955678] The account in question was found to be unresponsive to all five recall notifications. Additional information will be provided once the investigation has been completed.
Patient Sequence No: 1, Text Type: N, H10


[29874963] The reported device was not received for investigation; therefore, the reported failure cannot be confirmed. The complaint will be closed without a detailed investigation report and based on probable root cause. In the event that the device is received, the complaint will be reopened, a full evaluation will be conducted, and the investigation will be updated with the new results. Probable root cause for the reported failure involving this device could have been caused by: manufacturing nonconformity, nonconforming vesa mounting plate, weld fracture, damage during shipping /handling, missing or damaged fasteners/screws, use error. In sum, the reported failure could not be confirmed since the device was not received for investigation.
Patient Sequence No: 1, Text Type: N, H10


[29874964] It was reported that surgeon was hit on the shoulder and the nurse was hit in the face while trying to adjust the angle of the monitor.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0002936485-2014-00589
MDR Report Key4000392
Report Source07
Date Received2014-08-11
Date of Report2014-07-17
Date of Event2014-07-17
Date Mfgr Received2014-07-17
Date Added to Maude2014-08-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. THOMAS SHAFER
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal95138
Manufacturer Phone4087542000
Manufacturer G1STRYKER ENDOSCOPY-SAN JOSE
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal Code95138
Single Use3
Remedial ActionMA
Previous Use Code3
Removal Correction Number2936485-8/28/2013-002C
Event Type3
Type of Report3

Device Details

Brand NameSTRYKERVISION FLAT-PANEL MOUNTING ARM
Generic NameCART, EMERGENCY, CARDIOPULMONARY (EXCLUDING EQUIPMENT)
Product CodeBZN
Date Received2014-08-11
Catalog Number0240095200
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER ENDOSCOPY-SAN JOSE
Manufacturer Address5900 OPTICAL COURT SAN JOSE CA 95138 US 95138


Patients

Patient NumberTreatmentOutcomeDate
10 2014-08-11

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.