BIOGRAPH MMR 10433372

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2014-08-11 for BIOGRAPH MMR 10433372 manufactured by Siemens Healthcare.

Event Text Entries

[4706559] It was reported to siemens that while performing an mr pet examination, a female subject with large breasts complained of burns on her breast. It was reported that the spine coil, as well as two body matrix coils, were used during the examination. The left breast was observed to be warm to the touch, but not tender and no blister was present. The right breast was observed to be warm to the touch, tender to palpation and a small blister was present. The subject was treated with topical ointment. The study was approximately 2 hours in length.
Patient Sequence No: 1, Text Type: D, B5


[12308682] An investigation is currently underway. No system or coil malfunction has been found to date. Siemens will continue further investigation of the system coils to determine root cause. A supplement report will be submitted upon completion of the investigation.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2240869-2014-06205
MDR Report Key4003114
Report Source06
Date Received2014-08-11
Date of Report2014-07-24
Date of Event2014-07-23
Date Mfgr Received2014-07-24
Device Manufacturer Date2011-06-06
Date Added to Maude2014-08-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMEREDITH ADAMS
Manufacturer Street51 VALLEY STREAM PARKWAY, D02
Manufacturer CityMALVERN PA 19355
Manufacturer CountryUS
Manufacturer Postal19355
Manufacturer Phone6104483237
Manufacturer G1SIEMENS AG
Single Use3
Remedial ActionIN
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBIOGRAPH MMR
Generic NameTOMO IMAGER WITH NUC. MAG. RES
Product CodeOUO
Date Received2014-08-11
Model Number10433372
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE
Manufacturer Address51 VALLEY STREAM PARKWAY, D02 MALVIN PA 19355 US 19355


Patients

Patient NumberTreatmentOutcomeDate
10 2014-08-11