BD FACS SAMPLE PREP ASSISTANT III (BD SPA III 647205

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2014-08-08 for BD FACS SAMPLE PREP ASSISTANT III (BD SPA III 647205 manufactured by Tecan Systems Inc.

Event Text Entries

[4709888] On (b)(6) 2014 a (b)(6) bd field service engineer (fse) at (b)(6), dept of haematology, (b)(6) was exposed to biohazard waste during a service visit. Reported to bd biosciences on (b)(6) 2014. The exit waste quick-connect on the fluidics tower of the bd facs spa iii was broken and the engineer's visit was to resolve the situation. The male part of the fitting was jammed in the female part. The engineer proceeded to replace the female quick-disconnect mounted in the fluidics tower. Decontamination of the system by the customer was not possible due to the failure mode. Upon cutting the tubing to the fluidic fitting, residual pressure in the waste tube sprayed biohazardous waste at the engineer's face. The engineer was immediately taken to the accident and emergency dept (er) for a hepatitis b injection and immediate hiv post-exposure prophylaxis (pep) and treatment continues.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2916837-2014-00001
MDR Report Key4006471
Report Source07
Date Received2014-08-08
Date of Report2014-08-08
Date of Event2014-07-10
Date Mfgr Received2014-07-17
Device Manufacturer Date2008-05-27
Date Added to Maude2014-08-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA0
Event Location0
Manufacturer ContactERIC CLAUSSEN, DIR, REG COMPL
Manufacturer Street2350 QUME DR.
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4089546348
Single Use3
Remedial ActionRP
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameBD FACS SAMPLE PREP ASSISTANT III (BD SPA III
Generic NameBD SPA III
Product CodePER
Date Received2014-08-08
Model Number647205
Catalog Number647205
Lot NumberNA
ID NumberNA
OperatorOTHER
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTECAN SYSTEMS INC
Manufacturer Address2450 ZANKER RD SAN JOSE CA 95131 US 95131


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2014-08-08