MAUDE MDR 4013943

MDR report key
4013943
Report number
2024601-2014-00375
Event key
0
Event type
3
Date of event
2014-05-19
Date received
2014-07-16
Adverse event
3
Product problem
3
Patients in event
0
Reporter occupation
1
Health professional
3
Initial report to FDA
3
Event location
3

Manufacturer Contact#

Contact
LAURA LEBOEUF
Address
1120 S CAPITAL OF TEXAS HWY BULDING 1, SUITE 300 AUSTIN TX 78748 US
Phone
855-855-8555
Report source
M
Manufacturer link flag
Y

Devices#

Seq, Brand, Generic table
SeqBrandGenericManufacturerProduct codeModelCatalogLotPMA510(k)ImplantEvaluatedAvailability
1LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM 9.75CMIMPLANT, INTRAGASTRIC FOR MORBID OBESITYALLERGANLTINAB-20210S98LB019Y R

Patients#

Sequence, Received, Treatment table
SequenceReceivedTreatmentOutcome
12014-07-160

Event Narratives#

D

Patient 1

HEALTHCARE PROFESSIONAL REPORTS "LEAK AT LEVEL OF ACCESS PORT."

N

Patient 1

THE PRODUCT ASSOCIATED WITH THIS REPORT HAS NOT YET BEEN RETURNED. BASED UPON THE SERIAL NUMBER AND IMPLANT DATE THE CONNECTOR TYPE IS ASSUMED TO BE A STRAIN RELIEF. THE REPORTED OF THE EVENT WAS ASKED TO RETURN THE PRODUCT FOR ANALYSIS. VISUAL EXAMINATION MAY CONFIRM OR DETERMINE ANOTHER CONNECTOR TYPE ASSOCIATED WITH THIS REPORT. THE REPORTER OF THE EVENT WAS ASKED TO RETURN THE PRODUCT FOR ANALYSIS. ALLERGAN HAS NOT RECEIVED THE PRODUCT AT THIS TIME; THEREFORE, NO ANALYSIS OR TESTING HAS BEEN DONE. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OR THE CONNECTOR TUBING."

N

Patient 1

TAPER I. VISUAL EXAMINATION OF THE RETURNED DEVICE DETERMINED THE ACCESS PORT TUBING CONNECTOR TO BE A TAPER I. THE BAND TUBING WAS SEPARATED BROKEN AND HAD A SHARP UNIDENTIFIED OPENING. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "CAUTION: FAILURE TO CREATE A STABLE, SMOOTH PATH FOR THE ACCESS PORT TUBING, WITHOUT SHARP TURNS OR BENDS, CAN RESULT IN TUBING BREAKS AND LEAKAGE." "CAUTION: CARE MUST BE TAKEN DURING BAND ADJUSTMENT TO AVOID PUNCTURING THE TUBING WHICH CONNECTS THE ACCESS PORT AND BAND, AS THIS WILL CAUSE LEAKAGE AND DEFLATION OF THE INFLATABLE SECTION."

D

Patient 1

HEALTHCARE PROFESSIONAL REPORTS "LEAK AT LEVEL OF ACCESS PORT."