TQ-PREP WORKSTATION 6605429

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2014-08-19 for TQ-PREP WORKSTATION 6605429 manufactured by Beckman Coulter.

Event Text Entries

[4877167] The customer reported low dispensing volume on the first tube prepped when using the tq-prep workstation. The tq-prep was delivering low volumes of reagents a (lyse) and b (quench) on the first tube prepped after powering up the unit. A beckman coulter field service engineer (fse) was sent to the customer's facility to evaluate the system. There were no erroneous test results with this event. There was no death, injury or affect to user or patient treatment.
Patient Sequence No: 1, Text Type: D, B5


[12342676] (b)(6). On (b)(4) 2014, a beckman coulter field service engineer (fse) evaluated the system and replaced the syringes delivering reagents a and b in the instrument resolving the problem. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1061932-2014-01984
MDR Report Key4021681
Report Source01,05,06
Date Received2014-08-19
Date of Report2014-07-24
Date of Event2014-07-24
Date Mfgr Received2014-07-24
Device Manufacturer Date2011-03-15
Date Added to Maude2014-09-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER
Manufacturer Street11800 SW 147TH AVENUE
Manufacturer CityMIAMI FL 33196
Manufacturer CountryUS
Manufacturer Postal Code33196
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameTQ-PREP WORKSTATION
Generic NameSTATION, PIPETTING AND DILUTING, FOR CLINICAL USE
Product CodeJQW
Date Received2014-08-19
Model NumberNA
Catalog Number6605429
Lot NumberNA
ID NumberSOFTWARE VERSION: 1.1
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER
Manufacturer Address250 S. KRAEMER BLVD. BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2014-08-19

© 2024 FDA.report