ICON 25HG (COMBO CASSETTE) FHC202

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,08 report with the FDA on 2014-08-15 for ICON 25HG (COMBO CASSETTE) FHC202 manufactured by Alere San Diego, Inc..

Event Text Entries

[4724507] Caller alleging receiving a false negative on icon 25 hcg, one lot/patient, urine sample vs a positive on serum quant. Full detailed description: on (b)(6) 2014, a patient came in to physician's office to have iud placement. The physician did a urine pregnancy test, and the result came out negative. Customer stated that control line was good and that they waited a full 3 minute-read time and they had a second person verify the result. No retest was done on this test at that time. About 4 weeks later, patient complaint of discomfort. A quantitative serum test was done an the result came out positive (+) at 163,00 miu/ml, which, according to the customer, is equivalent to about 8 weeks. The lot number of the icon 25 used is l/n (b)(4), exp'n date: (b)(4) 2015. No patient sample was saved because their policy is to keep samples for only 48 hours. On (b)(4) 2014 customer reported that the physician removed the patient's iud and patient is fine. No patient information was provided. No further information available.
Patient Sequence No: 1, Text Type: D, B5


[12276363] Investigation pending.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2027969-2014-00758
MDR Report Key4036661
Report Source05,08
Date Received2014-08-15
Date of Report2014-07-28
Date of Event2014-06-19
Date Mfgr Received2014-07-28
Date Added to Maude2014-09-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactYA-LING KING
Manufacturer Street9975 SUMMERS RIDGE RD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052084
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameICON 25HG (COMBO CASSETTE)
Generic NameHCG PREGNANCY
Product CodeDKZ
Date Received2014-08-15
Model NumberFHC202
Lot NumberHCG3100047
Device Expiration Date2015-09-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC.
Manufacturer AddressSAN DIEGO CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2014-08-15

© 2026 FDA.report