MILTENYI CRYOMACS FREEZING BAG 500 FREEZE BAG 500 200-074-402

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2014-09-03 for MILTENYI CRYOMACS FREEZING BAG 500 FREEZE BAG 500 200-074-402 manufactured by Miltenyi Biotec.

Event Text Entries

[4730254] Infusion of cryopreserved autologous hpc, apheresis cellular therapy product (ctp) was scheduled for (b)(6) 2014. Three cryopreserved product bags were released from the stem cell lab and distributed to the pt bedside for infusion. Thaw and admin of the first and third units was performed without incident. Upon thaw of the second unit, leakage of the product into the secondary thawing bag was noted by the technologist performing the thaw. The technologist immediately identified the source of the leak and clamped the affected area. The leakage occurred at the base of the attached tubing (utilized for dispensing product/reagents into the cryo bag) just below where the tubing is heat sealed by the technologists at the completion of precessing. The integrity failure occurred at this location of the tubing where the tubing welds to the parent container and broke off at this area of the tubing. A 10 ml of product was lost as a result of the leakage. Due to medical necessity, it was determined by medical staff to infuse the product. Blood cultures were drawn on the pt and finalized on (b)(6) 2014 as "no growth". Post-thaw sterility samples were obtained from the affected product and sent for sterility testing. The post-thaw sterility testing was finalizing on (b)(6) 2014 as "no growth 14 days". Despite the product leakage, the pt received an optimal cd34+ cell dose. Per the pt operative report, the pt tolerated the procedure well and there were no expected and/or unexpected adverse reactions related to the product infusion.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number4055724
MDR Report Key4055724
Date Received2014-09-03
Date of Report2014-08-07
Date of Event2014-07-25
Date Facility Aware2014-07-25
Report Date2014-08-11
Date Reported to Mfgr2014-08-07
Date Added to Maude2014-09-05
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameMILTENYI CRYOMACS FREEZING BAG 500
Generic NameCRYOPRESERVATION FREEZE BAG
Product CodeKSR
Date Received2014-09-03
Model NumberFREEZE BAG 500
Catalog Number200-074-402
Lot Number6130206003
Device Expiration Date2016-02-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age4 MO
Device Sequence No1
Device Event Key0
ManufacturerMILTENYI BIOTEC
Manufacturer AddressBERGISCH GLADBACH GM


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2014-09-03