PURCEPTION 80 PERCENT LOWER PHASE GRANDIENT ART-2080

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06,08 report with the FDA on 2014-08-06 for PURCEPTION 80 PERCENT LOWER PHASE GRANDIENT ART-2080 manufactured by Sage In-vitro Fertilization Inc..

Event Text Entries

[4811584] "we have received complaint from the israeli health ministry that in one of our sperm laboratory unit, three women were insemination (iui) from sperm donations and almost immediately after injection (after half an hour) developed intense pain in the lower abdomen and leukocytes from 14 to 18 thousand, the sperm donations were from different donors, there were no similar cases in the past, the injection was prepared by different employees and medical staff believes that due to the rapid development of symptoms it seems to them that the cause is not a bacterial infection but rather atoxic response. "
Patient Sequence No: 1, Text Type: D, B5


[12228124] (b)(4). Conclusion: an assignable caused could not be determined as the products met all approved release specification at the time of manufacture before they were released. Based on review of all the information gathered during this investigation-associated sage documentation, manufacturing process, complaint history, and inspection of current inventory/retention sample, several external sources may have contributed to the reported complaint condition. Potential sources of product contamination include re-staging activities prior to procedure, the catheter used, or improper washing of the sperm samples (see fishbone diagram above). The retention samples of the subject lots tested show that the product meets all approved release specifications. The customer did not return the actual samples involved. The product performs as intended. No further actions are needed at this time. The investigation may be reevaluated with any new additional information received.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2031459-2014-00002
MDR Report Key4058347
Report Source01,05,06,08
Date Received2014-08-06
Date of Report2014-07-15
Date of Event2014-07-15
Date Mfgr Received2015-07-15
Device Manufacturer Date2013-07-16
Date Added to Maude2014-09-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactNANA BANAFO
Manufacturer Street75 CORPORATE DR
Manufacturer CityTRUMBULL CT 06611
Manufacturer CountryUS
Manufacturer Postal06611
Manufacturer Phone2036015200
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePURCEPTION 80 PERCENT LOWER PHASE GRANDIENT
Generic NamePURECEPTION MEDIA
Product CodeMQL
Date Received2014-08-06
Catalog NumberART-2080
Lot NumberD197 C
Device Expiration Date2015-07-16
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSAGE IN-VITRO FERTILIZATION INC.
Manufacturer AddressPASADENA CA US


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2014-08-06