STERILE DISPOSABLE TREPHINE BLADE 9.0MM 0009717

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2002-07-22 for STERILE DISPOSABLE TREPHINE BLADE 9.0MM 0009717 manufactured by Medtronic Xomed, Inc..

Event Text Entries

[268783] Customer reported that box contained the wrong size trephine blades-10mm blades instead of the 9mm that was printed on the product label. The customer returned the product in it's original unopened sterile pouches. The returned product inside of the boxes labeled 0009717 were actually part number 0009721, lot number 25214300.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1045254-2002-00012
MDR Report Key405862
Report Source06
Date Received2002-07-22
Date of Report2002-07-19
Date of Event2002-06-10
Date Mfgr Received2002-06-10
Device Manufacturer Date2002-05-01
Date Added to Maude2002-07-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactBILL WIGGINS
Manufacturer Street6743 SOUTHPOINT DR. N.
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal32216
Manufacturer Phone9043328196
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Remedial ActionRC
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTERILE DISPOSABLE TREPHINE BLADE 9.0MM
Generic NameDISPOSABLE OPHTHALMIC INSTRUMENT
Product CodeHRH
Date Received2002-07-22
Returned To Mfg2002-06-19
Model Number0009717
Catalog Number0009717
Lot Number25209500
ID Number*
Device Expiration Date2010-05-01
Device AvailabilityR
Device Eval'ed by MfgrY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key394851
ManufacturerMEDTRONIC XOMED, INC.
Manufacturer Address6743 SOUTHPOINT DR. NORTH JACKSONVILLE FL 32216 US
Baseline Brand NameSTERILE DISPOSABLE TREPHINE BLADE 9.0MM
Baseline Generic NameDISPOSABLE OPHTHALMIC INSTRUMENT
Baseline Model No0009717
Baseline Catalog No0009717
Baseline ID*


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2002-07-22