DIMENSION? CLINICAL CHEMISTRY SYSTEM RF420

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2014-09-04 for DIMENSION? CLINICAL CHEMISTRY SYSTEM RF420 manufactured by Siemens Healthcare Diagnostics Inc - Glasgow.

Event Text Entries

[17612850] A customer observed biased low mass creatinine kinase mb isoenzyme (mmb) results on patient crossover and qc samples on lot ed5167 reagent cartridges. No patient results were reported to physicians. Patient treatment was not altered or prescribed on the basis of the biased low mmb results. There was no report of adverse health consequences as a result of the biased low mmb results.
Patient Sequence No: 1, Text Type: D, B5


[17808969] The cause of the biased low mass creatinine kinase mb isoenzyme (mmb) results is lot to lot variability of the mass creatine kinase mb isoenzyme flex? Reagent cartridge. The issue was resolved by the customer with recalibration which provided better correlation to the customer's alternate dimension instrument. Siemens healthcare technical service laboratory conducted an investigation of lot ed5167 and did not confirm the customer's complaint for lot to lot variability. No further investigation is required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2517506-2014-00184
MDR Report Key4059622
Report Source05,06
Date Received2014-09-04
Date of Report2014-08-06
Date of Event2014-08-06
Date Mfgr Received2014-08-06
Device Manufacturer Date2014-06-16
Date Added to Maude2014-09-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJAMES MORGERA
Manufacturer StreetGLASGOW BUSINESS COMMUNITY PO BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal197146101
Manufacturer Phone3026318356
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer StreetGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE
Manufacturer CityNEWARK DE 19714
Manufacturer CountryUS
Manufacturer Postal Code19714
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDIMENSION? CLINICAL CHEMISTRY SYSTEM
Generic NameMASS CREATINE KINASE MB ISOENZYME FLEX? REAGENT CARTRIDGE
Product CodeJHY
Date Received2014-09-04
Catalog NumberRF420
Lot NumberED5167
Device Expiration Date2015-06-16
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer AddressGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE NEWARK DE 19714 US 19714


Patients

Patient NumberTreatmentOutcomeDate
10 2014-09-04

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.