VIVIDIMAGE SURGICAL GRADE MONITOR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2014-09-05 for VIVIDIMAGE SURGICAL GRADE MONITOR manufactured by Vts Medical Systems, Llc.

Event Text Entries

[16185397] The user facility reported that during a patient procedure the monitor signal was going in and out. No injuries or procedural delays/cancellations were reported.
Patient Sequence No: 1, Text Type: D, B5


[16443411] A steris service technician inspected the monitor and found a loose wire to the vga (video graphics array) connection. The technician reseated and secured the wire to the vga connection. The monitor was returned to service and no further issues have been reported.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1000404456-2014-00006
MDR Report Key4064296
Report Source05
Date Received2014-09-05
Date of Report2014-09-05
Date of Event2014-08-08
Date Mfgr Received2014-08-08
Date Added to Maude2014-09-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. KATHRYN CADORETTE
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927231
Manufacturer G1VTS MEDICAL SYSTEMS, LLC
Manufacturer Street40 MELVILLE PARK RD
Manufacturer CityMELVILLE NY 11747
Manufacturer CountryUS
Manufacturer Postal Code11747
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVIVIDIMAGE SURGICAL GRADE MONITOR
Generic NameMONITOR
Product CodeKQM
Date Received2014-09-05
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerVTS MEDICAL SYSTEMS, LLC
Manufacturer Address40 MELVILLE PARK RD MELVILLE NY 11747 US 11747


Patients

Patient NumberTreatmentOutcomeDate
10 2014-09-05

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.