MAESTRO MINI-TWIN BRACKET WITH HOOK 711-399

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 99 report with the FDA on 2014-08-12 for MAESTRO MINI-TWIN BRACKET WITH HOOK 711-399 manufactured by Ortho Organizers, Inc..

Event Text Entries

[18935014] On (b)(6) 2013, brackets were placed. Initial diagnosis: class i 18 months, moderate crowding. Tooth #11 perfectly healthy with no signs of previous restorations, decay, or decalcification. Dentist was removing bracket from ul3 (tooth #11) and tooth broke horizontally in half. Post incident diagnosis: #11 horizontal fracture with small pulpal involvement. No signs of decalcification or decay. Clean fracture with bracket still attached to the fractured portion. 2 lugs broken during removal. Tooth #11 will need endo with build up and emax crown at a later date. Tooth was bonded back together and new bracket placed on tooth. (b)(6).
Patient Sequence No: 1, Text Type: D, B5


[19134252] Tooth fracture can occur upon debonding (removal) of bracket. Although most fractures are limited to enamel, the potential for pulpal involvement exists. Literature describing debonding adverse events can be tracked back to at least 1996. Providers are trained to use care when debonding brackets.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2081322-2014-00001
MDR Report Key4068001
Report Source99
Date Received2014-08-12
Date of Report2014-08-12
Date of Event2014-07-15
Date Mfgr Received2014-07-16
Date Added to Maude2014-09-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street1822 ASTON AVENUE
Manufacturer CityCARLSBAD CA 92008
Manufacturer CountryUS
Manufacturer Postal92008
Manufacturer Phone7604488600
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMAESTRO MINI-TWIN BRACKET WITH HOOK
Product CodeEJF
Date Received2014-08-12
Model Number711-399
Catalog Number711-399
Lot NumberUNK
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerORTHO ORGANIZERS, INC.
Manufacturer Address1822 ASTON AVENUE CARLSBAD CA 92008760 US 92008 7603


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2014-08-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.