COULTER DXH DILUENT 628017

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2014-09-08 for COULTER DXH DILUENT 628017 manufactured by Beckman Coulter.

Event Text Entries

[4897073] The customer reported platelet background high failures on the unicel dxh 800 coulter cellular analysis system when using a specific lot number of coulter dxh diluent. There were no erroneous test results with this event. There was no death, injury or affect to user or patient treatment.
Patient Sequence No: 1, Text Type: D, B5


[12428399] The customer replaced coulter dxh diluent lot number 3510540 with coulter dxh diluent lot number 3510400, primed the diluent, and the platelet background counts were acceptable. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1061932-2014-02292
MDR Report Key4070043
Report Source05,06
Date Received2014-09-08
Date of Report2014-08-14
Date of Event2014-08-14
Date Mfgr Received2014-08-14
Device Manufacturer Date2014-06-23
Date Added to Maude2014-09-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER
Manufacturer Street11800 SW 147TH AVENUE
Manufacturer CityMIAMI FL 33196
Manufacturer CountryUS
Manufacturer Postal Code33196
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction Number2050012-08/25/2014-003C
Event Type3
Type of Report3

Device Details

Brand NameCOULTER DXH DILUENT
Generic NameBLOOD CELL DILUENT
Product CodeGIF
Date Received2014-09-08
Model NumberNA
Catalog Number628017
Lot Number3510540
Device Expiration Date2015-12-22
OperatorHEALTH PROFESSIONAL
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER
Manufacturer Address250 S. KRAEMER BLVD. BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2014-09-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.