IC PROFILER-MR 1123

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2014-09-03 for IC PROFILER-MR 1123 manufactured by Sun Nuclear Corp..

Event Text Entries

[16310032] Staff using an ic profiler-mr at (b)(6) radiation oncology reported their ic profiler-mr unit that was resting on the couch moved when the couch was translated into a magnetic field of a viewray delivery system. No injury was reported.
Patient Sequence No: 1, Text Type: D, B5


[16385643] The ic profiler-mr is not used in the presence of a patient. However, staff setting up the ic profiler-mr for measurement of an installed viewray delivery system could experience a pinch hazard between the covers on the delivery system and the ic profiler-mr if it is attracted towards the delivery system by the magnetic field. Investigation identified an assembly error of this unit. There were 3 units manufactured in the lot and 9 units manufactured in total. As part of the investigation of this incident, all units manufactured have been evaluated. Data obtained confirmed 2 units contained the assembly error. These units have been removed and repaired.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1038814-2014-00001
MDR Report Key4084960
Report Source06
Date Received2014-09-03
Date of Report2014-08-06
Date of Event2014-08-06
Date Mfgr Received2014-08-06
Device Manufacturer Date2014-06-01
Date Added to Maude2014-09-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJONATHAN LEE
Manufacturer Street3275 SUNTREE BLVD
Manufacturer CityMELBOURNE FL 32940
Manufacturer CountryUS
Manufacturer Postal32940
Manufacturer Phone3212596862
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIC PROFILER-MR
Generic NameIC PROFILER-MR
Product CodeLHO
Date Received2014-09-03
Returned To Mfg2014-08-08
Model Number1123
Lot NumberW92327
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSUN NUCLEAR CORP.
Manufacturer AddressMELBOURNE FL US


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2014-09-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.