OLYMPUS HF RESECTION ELECTRODES A22206C

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,06,07 report with the FDA on 2014-09-05 for OLYMPUS HF RESECTION ELECTRODES A22206C manufactured by Olympus Winter & Ibe Gmbh.

Event Text Entries

[21724461] Olympus was informed that during a transurethral resection of the prostate (turp) procedure, the loop on the device broke off inside the pt. The procedure was completed with a second electrode. No additional info was provided.
Patient Sequence No: 1, Text Type: D, B5


[22144616] This supplemental report is being submitted to provide additional info based on the eval of the device. The eval confirmed that the electrode was damaged and broken. It was noted that the loop wire had broken off at both sides of the red filter section, and there was evidence of thermal damage and melting at the damaged areas. The device has been forwarded to the original equipment mfr for further investigation. If additional info is available at a later time, this report will be supplemented accordingly.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2951238-2014-00422
MDR Report Key4098928
Report Source01,06,07
Date Received2014-09-05
Date of Report2014-05-23
Date of Event2014-05-01
Date Mfgr Received2014-05-23
Date Added to Maude2014-09-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactNOEMI SCHAMBACH
Manufacturer Street2400 RINGWOOD AVE.
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4089355002
Manufacturer G1OLYMPUS WINTER & IBE GMBH
Manufacturer StreetKUEHNSTRASSE 61
Manufacturer CityHAMBURG 22045
Manufacturer CountryGM
Manufacturer Postal Code22045
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS HF RESECTION ELECTRODES
Generic NameRESECTION ELECTRODES
Product CodeFDC
Date Received2014-09-05
Returned To Mfg2014-07-16
Model NumberA22206C
Catalog NumberA22206C
Lot Number13282P02L001
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS WINTER & IBE GMBH
Manufacturer AddressKUEHNSTRASSE 61 HANBURG 22045 GM 22045


Patients

Patient NumberTreatmentOutcomeDate
10 2014-09-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.