MCP SZ. 40 PROXIMAL WW MCP-100-40P-WW

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,07 report with the FDA on 2014-08-28 for MCP SZ. 40 PROXIMAL WW MCP-100-40P-WW manufactured by Ascension Orthopedics.

Event Text Entries

[4719494] It was reported during a two joint pyrocarbon mcp surgical case, the shaft of a size 40 proximal mcp joint fractured mid stern, in the intramedullary canal. It took extra time to remove the fractured implant. A spare device was available and used to complete the surgery. There was no injury to the patient reported. The surgery time was increased 20 minutes.
Patient Sequence No: 1, Text Type: D, B5


[12352571] The device involved in the reported incident has been received for evaluation. An investigation has been initiated based on the reported information.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1651501-2014-00043
MDR Report Key4102668
Report Source05,06,07
Date Received2014-08-28
Date of Report2014-08-01
Date of Event2014-08-01
Date Mfgr Received2014-08-01
Date Added to Maude2014-09-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCAREN FINKELSTEIN
Manufacturer Street315 ENTERPRISE DRIVE
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099362341
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMCP SZ. 40 PROXIMAL WW
Product CodeNEG
Date Received2014-08-28
Returned To Mfg2014-08-08
Catalog NumberMCP-100-40P-WW
Lot Number13-0086
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerASCENSION ORTHOPEDICS
Manufacturer AddressAUSTIN TX 78754 US 78754


Patients

Patient NumberTreatmentOutcomeDate
10 2014-08-28

© 2026 FDA.report