STATSPIN MP X00-003917-001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2014-08-28 for STATSPIN MP X00-003917-001 manufactured by Iris International.

Event Text Entries

[4758249] Customer reported a rotor breakage.
Patient Sequence No: 1, Text Type: D, B5


[12198080] A loud noise and rotor breakage was reported by the centrifuge. An infant pt sample was lost as a result leading to recollection of sample. It was reported that this delayed pt treatment; however, it is unk if there was any harm to the pt due to the delay. The pieces of rotor and pt sample were contained within the centrifuge and no injuries to user were reported.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2023446-2014-00137
MDR Report Key4102691
Report Source05,06
Date Received2014-08-28
Date of Report2013-07-29
Date of Event2014-07-29
Date Mfgr Received2014-07-29
Device Manufacturer Date2008-10-01
Date Added to Maude2014-09-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactSUDHA GUPTA
Manufacturer Street9172 ETON AVE.
Manufacturer CityCHATSWORTH CA 91311
Manufacturer CountryUS
Manufacturer Postal91311
Manufacturer Phone8185277272
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTATSPIN MP
Generic NameSTATSPIN CENTRIFUGE
Product CodeGKG
Date Received2014-08-28
Catalog NumberX00-003917-001
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerIRIS INTERNATIONAL
Manufacturer Address9172 ETON AVE. CHATSWORTH CA 91311 US 91311


Patients

Patient NumberTreatmentOutcomeDate
10 2014-08-28

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.