CELLEBRITY? M00516011 1601

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,07 report with the FDA on 2014-09-23 for CELLEBRITY? M00516011 1601 manufactured by Boston Scientific - Spencer.

Event Text Entries

[4922181] It was reported to boston scientific corporation that a celebrity cytology brush was used in the lungs during a lung biopsy procedure performed on (b)(6) 2014. According to the complainant, during the procedure, an excelon needle and a brush failed to obtain a tissue sample, the needle failed to flush and the brush failed to extend. The procedure was completed with another celebrity cytology brush. There were no patient complications reported as a result of this event. The patient's condition at the conclusion of the procedure was reported to be fine. This event has been deemed a reportable event based on the investigation results: catheter torn.
Patient Sequence No: 1, Text Type: D, B5


[12370457] Investigation results finding the catheter torn. Investigation results: visual evaluation of the returned device found the working length was bent and brush was retracted. Functional evaluation performed, the brush was attempted to be extended but failed due to the tear in the catheter, but the strain relief was still connected to the handle. The condition of the returned device was consistent with the complaint that the brush was difficult to extend; the complaint was confirmed. The proximal part of the catheter was torn but the strain relief remained intact. Therefore, a review and analysis of all available information indicated that the most probable root cause is operational context. A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.? A search of the complaint database confirmed that no other complaints exist for the specified batch.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3005099803-2014-03138
MDR Report Key4111811
Report Source05,07
Date Received2014-09-23
Date of Report2014-08-31
Date of Event2014-07-28
Date Mfgr Received2014-08-31
Device Manufacturer Date2014-04-07
Date Added to Maude2014-09-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactEMP. NANCY CUTINO
Manufacturer Street100 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834000
Manufacturer G1BOSTON SCIENTIFIC - SPENCER
Manufacturer Street780 BROOKSIDE DRIVE
Manufacturer CitySPENCER IN 47460
Manufacturer CountryUS
Manufacturer Postal Code47460
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCELLEBRITY?
Generic NameBRUSH, BIOPSY, BRONCHOSCOPE (NON-RIGID)
Product CodeBTG
Date Received2014-09-23
Returned To Mfg2014-08-01
Model NumberM00516011
Catalog Number1601
Lot Number16871287
Device Expiration Date2016-04-03
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC - SPENCER
Manufacturer Address780 BROOKSIDE DRIVE SPENCER IN 47460 US 47460


Patients

Patient NumberTreatmentOutcomeDate
10 2014-09-23

© 2024 FDA.report