STRATAPULE PORTABLE STRATAPULSE PORTABLE NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2014-09-15 for STRATAPULE PORTABLE STRATAPULSE PORTABLE NA manufactured by Shanghai Apolo Medical Tech Co Ltd..

Event Text Entries

[4922784] On (b)(6) 2013, mr (b)(4) was repairing this device for a different problem. During his analysis be discovered that the spot size was set to 15 x 50 when it should have been 12 x 35. On (b)(6) 2014, as part of a complaint remediation investigation, i asked him about this malfunction. He stated that it was his opinion that if this malfunction recurred it could result in too much energy being delivered to the treatment area and cause permanent injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005327046-2014-00003
MDR Report Key4115748
Date Received2014-09-15
Date of Report2014-09-12
Date Facility Aware2014-09-10
Report Date2014-09-12
Date Reported to Mfgr2014-09-12
Date Added to Maude2014-09-25
Event Key0
Report Source CodeDistributor report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street12585 OLD HIGHWAY 280 STE 103
Manufacturer CityCHELSEA AL 350430000
Manufacturer CountryUS
Manufacturer Postal350430000
Manufacturer G1MARKET TECHNOLOGIES INC
Manufacturer Street12585 OLD HIGHWAY 280 STE 103
Manufacturer CityCHELSEA AL 35043000
Manufacturer CountryUS
Manufacturer Postal Code35043 0000
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameSTRATAPULE PORTABLE
Generic NameIPL SYSTEM
Product CodeONF
Date Received2014-09-15
Model NumberSTRATAPULSE PORTABLE
Catalog NumberNA
Lot NumberNA
Device AvailabilityN
Device Age15 MO
Device Sequence No1
Device Event Key0
ManufacturerSHANGHAI APOLO MEDICAL TECH CO LTD.
Manufacturer Address3F, BLDG A, NO 388 YINDU RD XUHUI DISTRICT, SHANGHAI AL CH


Patients

Patient NumberTreatmentOutcomeDate
10 2014-09-15

© 2026 FDA.report