HALSTED-MOSQUITO FCPSDELSTR1X2125MM BH120R

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,07 report with the FDA on 2014-07-28 for HALSTED-MOSQUITO FCPSDELSTR1X2125MM BH120R manufactured by Aesculap Ag & Co. Kg.

Event Text Entries

[18369493] Country of complaint: (b)(6). Complaint: the mouse tooth of the jaw was missing. When the surgeon tried to use this product in surgery, e found that the mouse tooth of the jaw was missing. The fragment has not been found yet. Primary surgery: (b)(6) 2014. Revision surgery: none. Pt data: unk. Further detail: none. Involved component: none. Reason for vigilance suspicious: fragment has not been found.
Patient Sequence No: 1, Text Type: D, B5


[18465012] Us reporting agent notified on (b)(4) 2014. Mfg site evaluation: the issue has been caused by a mfg error, which was not detected in final inspection. All responsible personnel have been informed regarding this issue in order to ensure it is avoided in the future. No further action required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2916714-2014-00583
MDR Report Key4120804
Report Source01,07
Date Received2014-07-28
Date of Report2014-07-28
Date of Event2014-04-24
Date Mfgr Received2014-04-24
Device Manufacturer Date2012-01-01
Date Added to Maude2014-10-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMICHELLE LINK
Manufacturer Street615 LAMBERT POINTE DR
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3145515938
Manufacturer G1AESCULAP AG & CO KG
Manufacturer StreetPO BOX 40
Manufacturer CityTUTTLINGEN 78501
Manufacturer CountryGM
Manufacturer Postal Code78501
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHALSTED-MOSQUITO FCPSDELSTR1X2125MM
Generic NameVASCULAR CLAMP
Product CodeHRQ
Date Received2014-07-28
Model NumberBH120R
Catalog NumberBH120R
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerAESCULAP AG & CO. KG
Manufacturer AddressTUTTLINGEN, DE 78532 78532


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2014-07-28