INFRAVISION IR ILLUMINATOR 0220180521

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2014-09-26 for INFRAVISION IR ILLUMINATOR 0220180521 manufactured by Stryker Endoscopy-san Jose.

Event Text Entries

[4938596] It was reported that the unit sparked when plugged in.
Patient Sequence No: 1, Text Type: D, B5


[12205260] Additional information will be provided once the investigation has been completed.
Patient Sequence No: 1, Text Type: N, H10


[29876909] The product was returned for investigation and the reported failure mode was confirmed. The following was observed: the unit was due for calibration. This does confirm the alleged failure mode of sparking probable root cause: console ground failure. Current leakage too high. The product was returned for investigation and the reported failure mode was confirmed. The failure mode will be monitored for future reoccurrence.???????
Patient Sequence No: 1, Text Type: N, H10


[29876910] It was reported that the unit sparked when plugged in.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0002936485-2014-00758
MDR Report Key4123305
Report Source06
Date Received2014-09-26
Date of Report2014-09-02
Date of Event2014-09-02
Date Mfgr Received2014-09-02
Date Added to Maude2014-10-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. THOMAS SHAFER
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal95138
Manufacturer Phone4087542000
Manufacturer G1STRYKER ENDOSCOPY-SAN JOSE
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal Code95138
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINFRAVISION IR ILLUMINATOR
Generic NameCARRIER, SPONGE, ENDOSCOPIC
Product CodeFGS
Date Received2014-09-26
Returned To Mfg2014-09-02
Catalog Number0220180521
OperatorOTHER
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER ENDOSCOPY-SAN JOSE
Manufacturer Address5900 OPTICAL COURT SAN JOSE CA 95138 US 95138


Patients

Patient NumberTreatmentOutcomeDate
10 2014-09-26

© 2026 FDA.report