LIGHT FANTASTIC II TRACK MOUNTED DENTAL LIGHT LFTII

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2014-10-07 for LIGHT FANTASTIC II TRACK MOUNTED DENTAL LIGHT LFTII manufactured by Pelton & Crane.

Event Text Entries

[4910565] It was reported that a pelton & crane dental light had come apart at the arm assembly and fell down towards the ground. No one was in the room when the event took place.
Patient Sequence No: 1, Text Type: D, B5


[12359528] The (b)(4) distributor inspected the dental light at the doctors office and was told by the dental office, the light arm rotational stop mechanism had seized together around a year ago; however, the facility continued to use the dental light without having it repaired. The function of the rotational stop is to prevent the light from rotating more than 350 degrees. As a result, the light arm assembly had rotated numerous times past 360 degrees during use allowing the light arm assembly to unscrew from the light post assembly and fall. The (b)(4)distributor repaired the light and placed it back into service.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1017522-2014-00014
MDR Report Key4156551
Report Source08
Date Received2014-10-07
Date of Report2014-09-09
Date of Event2014-09-04
Date Mfgr Received2014-09-09
Device Manufacturer Date2002-01-01
Date Added to Maude2014-10-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactFRANK RAY, MGR
Manufacturer Street11727 FRUEHAUF DR.
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal28273
Manufacturer Phone7045877227
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLIGHT FANTASTIC II TRACK MOUNTED DENTAL LIGHT
Generic NameDENTAL OPERATING LIGHT
Product CodeEAZ
Date Received2014-10-07
Model NumberLFTII
Catalog NumberLFTII
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerPELTON & CRANE
Manufacturer Address11727 FRUEHAUF DR. CHARLOTTE NC 28273 US 28273


Patients

Patient NumberTreatmentOutcomeDate
10 2014-10-07

© 2026 FDA.report