MICROSCAN PANELS UNK

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2014-10-10 for MICROSCAN PANELS UNK manufactured by Siemens Healthcare Diagnostics Inc.

Event Text Entries

[5021348] It was reported that the barcode printer was skipping blank labels and the only way to get it to stop would be to cycle and power on the printer. The customer had to repair the bar code label and that is when the print order was affected and resulted in a mix up of labels. Customer indicated that barcode printer was not the cause of mixed up labels. Approximately 11 panels were affected with incorrect barcodes. It is unk how many incorrect results were initially reported to the physicians. When the customer discovered the mixed up, isolates were re-tested. The corrected results were issued and reported to the physician. It is unk when the label mix up started to occur. The specific info about the treatment was not reported.
Patient Sequence No: 1, Text Type: D, B5


[12420076] (b)(4). The customer indicated that the microscan barcode printer was not at fault for this issue. The mixed up of specimen issues in their facility have been resolved with the implementation of a new process to confirm labels before attaching to the panels. The model numbers of the affected 11 panels ere not reported. Calibration of the barcode printer was performed on (b)(6) 2014 and the barcode printer skipped blank labels was resolved.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2919016-2014-00040
MDR Report Key4173501
Report Source05
Date Received2014-10-10
Date of Report2014-09-25
Date of Event2014-09-25
Date Mfgr Received2014-09-25
Date Added to Maude2014-11-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJOSE UNTALAN
Manufacturer Street2040 ENTERPRISE BLVD.
Manufacturer CityWEST SACRAMENTO CA 95691
Manufacturer CountryUS
Manufacturer Postal95691
Manufacturer Phone9163743031
Single Use0
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMICROSCAN PANELS
Generic NameNONE
Product CodeLRG
Date Received2014-10-10
Model NumberUNK
Catalog NumberUNK
Lot NumberUNK
ID NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC
Manufacturer AddressWEST SACRAMENTO CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2014-10-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.