WATER PURIFICATION SYSTEM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2014-10-17 for WATER PURIFICATION SYSTEM manufactured by Mar Cor Purification.

Event Text Entries

[18035286] The reverse osmosis loop's pretreatment dealkalizer scavenger head was not functioning correctly which caused the cwp ro unit to alarm due to high conductivity. Patients had treatment delayed.
Patient Sequence No: 1, Text Type: D, B5


[18392077] The facility reported they experienced a high conductivity alarm on their cwp ro system (percent rejection has dropped to (b)(4)). Upon inspection, the clinic found the reverse osmosis loop's pretreatment dealkalizer scavenger head was not functioning correctly. This allowed the cwp ro to intake some brine causing the high conductivity alarm and ro water production to shut down. The scavenger head was bypassed and the water conductivity began to fall. 16 patents were treatment delayed. There were no reports of patient illness or injury as a result of this incident. This complaint will continue to be maintained within mar cor purification's complaint system.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3019131-2014-00028
MDR Report Key4180216
Report Source05,06
Date Received2014-10-17
Date of Report2014-10-16
Date of Event2014-09-19
Date Mfgr Received2014-09-22
Device Manufacturer Date2014-05-26
Date Added to Maude2014-10-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMEGAN DICKEY
Manufacturer Street14550 28TH AVE N
Manufacturer CityMINNEAPOLIS MN 55447
Manufacturer CountryUS
Manufacturer Postal55447
Manufacturer Phone7635533300
Manufacturer G1MAR COR PURIFICATION
Manufacturer Street14550 28TH AVE N
Manufacturer CityMINNEAPOLIS MN 55447
Manufacturer CountryUS
Manufacturer Postal Code55447
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWATER PURIFICATION SYSTEM
Generic NameREVERSE OSMOSIS SYSTEM
Product CodeFIP
Date Received2014-10-17
ID Number3028501
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMAR COR PURIFICATION
Manufacturer Address14550 28TH AVE N MINNEAPOLIS MN 55447 US 55447


Patients

Patient NumberTreatmentOutcomeDate
10 2014-10-17

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.