[20328302] It was reported that smoke and sparks started to come out of the safety station. Additional follow up with the customer identified that the issue was observed at the time two of the maintenance employees were moving the unit away from the wall. While moving the unit, water dripped onto the service which led to the customer's reported issue. After the sparking occurred, the power was disconnected and the unit was taken to be examined. It was noted that the initially reported smoke may have been related to steam coming from water contact on the device. It was confirmed that at no point did sparking make contact with anyone handling the unit. There was no harm associated with the event.
Patient Sequence No: 1, Text Type: D, B5


[20566646] The device was not returned to the manufacturer, but reviewed by service technician at the user facility. A follow up medwatch will be submitted once the investigation is complete.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1954182-2014-00001
MDR Report Key4202520
Report Source06,07
Date Received2014-09-18
Date of Report2014-08-21
Date of Event2014-08-21
Date Mfgr Received2014-08-21
Date Added to Maude2014-10-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKATHLEEN SMITH
Manufacturer Street200 NW PKWY. RD.
Manufacturer CityRIVERSIDE MO 64150
Manufacturer CountryUS
Manufacturer Postal64150
Manufacturer Phone3303649447
Single Use0
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSAFETY STATION
Generic NameSAFETY STATION
Product CodeFLH
Date Received2014-09-18
Model NumberNA
Catalog NumberTP-SS100
Lot NumberNA
OperatorOTHER
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerDORNOCH MEDICAL
Manufacturer Address200 NW PKWY. RD. RIVERSIDE MO 64150 US 64150


Patients

Patient NumberTreatmentOutcomeDate
10 2014-09-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.