PILLCAM COLON2 CAPSULE FGS-0286

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2014-10-22 for PILLCAM COLON2 CAPSULE FGS-0286 manufactured by Given Imaging Ltd..

Event Text Entries

[5074722] On (b)(6) the vumc (b)(6) performed a vce c2 procedure. After connecting the capsule with the dr3 and ingestion the capsule the image have been seen on the display of the dr3. After approx 5 minutes the dr did not receive images anymore. Capsule signal was no longer detected. The hospital changed array and dr without success and possibility was mentioned that the problem was the capsule. Four days later the patient found the capsule in her stool and the dome was separated from the body of the capsule. The patient suffered no ill effects and remains completely asymptomatic.
Patient Sequence No: 1, Text Type: D, B5


[12443639] Upon receiving report, the customer was asked to return the capsule as soon as possible to help determine the cause of this malfunction. Unfortunately, the capsule has not been received one month after the request was made. In the interim, the following investigations were completed in the absence of the actual capsule involved: destructive test (test performed routinely to verify integrity of capsule assembly procedure including gluing process) for 23812c lot performed on (b)(4) 2013 and passed successfully. The dhr of this lot reviewed and no non conformity was found. No similar reports for the other 24 capsules from the same lot. The capsule gluing device and the capsule gluing processes validation reviewed and results showed complete conformity was found. The capsule component (dome and case) incoming inspection reviewed and meet the part specifications. The gluing coc was reviewed and conformity was found.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9710107-2014-00194
MDR Report Key4217041
Report Source06
Date Received2014-10-22
Date of Report2014-10-23
Date of Event2014-09-15
Device Manufacturer Date2013-12-01
Date Added to Maude2014-11-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePILLCAM COLON2 CAPSULE
Generic NameCOLON 2 CAPSULE
Product CodePGD
Date Received2014-10-22
Catalog NumberFGS-0286
Lot Number23812C
Device Expiration Date2014-12-02
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerGIVEN IMAGING LTD.
Manufacturer AddressHERMON BUILDING NEW INDUSTRIAL PARK, P.O.BOX 258 YOKNEAM 20692 IS 20692


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2014-10-22