CORPAK BLD-14660

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2002-10-09 for CORPAK BLD-14660 manufactured by Corpak Medsystems.

Event Text Entries

[257046] Over past 4-5 months, about 1 out of every 40-60 adult transport circuits breaks at the exhalation valve. The exhalation valve tubing port breaks off and breaks into one or more pieces, leaving a hole in the top cap of the valve about the size of 2/3 of a dime coin. This has occurred when mild to moderate pressure is applied in checking for secure connections of component parts prior to product use. Customer reports no prior experience of this event in several years use of this product. This event could cause a pt to suffer an untoward outcome of death or serious injury if it caused loss of ventilation to the pt, if this loss was not recognized by the caretaker in a timely fashion and/or if it occurred in a setting in which immediate other manual or mechanical ventilatory support was not available to the clinician and pt.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1419949-2002-00006
MDR Report Key423014
Report Source05,06
Date Received2002-10-09
Date Added to Maude2002-10-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationRESPIRATORY THERAPIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactSTEPHANIE WASIELEWSKI
Manufacturer Street100 CHADDICK DRIVE
Manufacturer CityWHEELING IL 60090
Manufacturer CountryUS
Manufacturer Postal60090
Manufacturer Phone8475374601
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCORPAK
Generic NameTRANSPORT VENTILATOR CIRCUIT
Product CodeCBP
Date Received2002-10-09
Returned To Mfg2002-10-02
Model NumberNA
Catalog NumberBLD-14660
Lot NumberUNK
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key412068
ManufacturerCORPAK MEDSYSTEMS
Manufacturer Address100 CHADDICK DR WHEELING IL 60090 US
Baseline Brand NamePULMANEX TRANSPORT CIRCUIT
Baseline Generic NameTRANSPORT CIRCUIT
Baseline Model NoNA
Baseline Catalog NoBLD-14660
Baseline Device FamilyTRANSPORT CIRCUIT
Baseline Shelf Life ContainedN
Baseline Shelf Life [Months]NA
Baseline PMA FlagN
Baseline 510K PMNN
Baseline PreamendmentN
Baseline TransitionalN
510k ExemptY


Patients

Patient NumberTreatmentOutcomeDate
10 2002-10-09

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.