MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06,07,company representative, report with the FDA on 2014-11-13 for PULSAR 2 GENERATOR PS100-102 manufactured by Medtronic Advanced Energy.
[5049568]
During a pacemaker generator change out procedure, the physician cut through a pacemaker lead with the plasmablade. The physician reported that he was trying to free the leads in the pacemaker pocket when he saw a spark and realized he had cut through a lead. Damaged lead was replaced with no patient impact. Pulsar system was on settings cut 5, coag 6. The lead was an attain 4196 lv.
Patient Sequence No: 1, Text Type: D, B5
[12560873]
Product event # (b)(4). Evaluation, (method), (result), (conclusion): generator still in use at facility therefore analysis unable to be performed. (b)(4).
Patient Sequence No: 1, Text Type: N, H10
[103813526]
If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1226420-2014-00108 |
MDR Report Key | 4247432 |
Report Source | 06,07,COMPANY REPRESENTATIVE, |
Date Received | 2014-11-13 |
Date of Report | 2014-10-16 |
Date of Event | 2014-10-16 |
Date Mfgr Received | 2014-10-16 |
Date Added to Maude | 2014-11-25 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | KAYLEE BOISVERT |
Manufacturer Street | 180 INTERNATIONAL DRIVE |
Manufacturer City | PORTSMOUTH NH 03801 |
Manufacturer Country | US |
Manufacturer Postal | 03801 |
Manufacturer Phone | 6038426234 |
Manufacturer G1 | MEDTRONIC ADVANCED ENERGY, LLC |
Manufacturer Street | 180 INTERNATIONAL DRIVE |
Manufacturer City | PORTSMOUTH NH 03801 |
Manufacturer Country | US |
Manufacturer Postal Code | 03801 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | PULSAR 2 GENERATOR |
Generic Name | GENERATOR,ELECTROSURGICAL,COAGULATION,CANCER |
Product Code | MUL |
Date Received | 2014-11-13 |
Model Number | PS100-102 |
Catalog Number | PS100-102 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MEDTRONIC ADVANCED ENERGY |
Manufacturer Address | 180 INTERNATIONAL DRIVE PORTSMOUTH NH 03801 US 03801 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2014-11-13 |